Water Purification System — Technical Specification

1. Water Quality Parameters

Parameter Type 1 (Ultrapure) Type 2 (Pure) Test Method
Resistivity 18.2 MΩ·cm >10 MΩ·cm Online monitor
TOC <5 ppb <50 ppb Online TOC analyzer
Bacteria <0.1 CFU/mL <10 CFU/mL Membrane filtration
Endotoxin <0.001 EU/mL <0.25 EU/mL LAL test
Particles (0.22 µm) <1/mL Membrane retention
Silica <0.01 ppb <0.1 ppb ICP-MS

2. System Components

Component Specification
Pre-treatment Sediment filter, carbon filter, water softener
Primary Purification Reverse osmosis (RO) membrane
Polishing Mixed-bed ion exchange, UV oxidation (185/254 nm)
Final Filtration 0.22 µm PES membrane filter
Storage 100 L polypropylene tank with UV sanitization

3. Performance

Parameter Value
Production Rate (RO) 10 – 20 L/h
Dispense Flow Rate Up to 2 L/min
Recovery Rate (RO) 50 – 75%
System Pressure 4 – 6 bar