Karl Fischer Titrator — Technical Specification

1. Technical Parameters

Parameter Specification
Method Coulometric (iodine generation)
Measuring Range 0.1 µg – 100 mg H₂O
Resolution 0.1 µg H₂O
Accuracy ±3 µg H₂O (at <100 µg)
Titration Current Auto-controlled
Stirring Magnetic, adjustable speed
Cell Type Diaphragm or diaphragmless
Anode Solution Hydranal Coulomat AG
Cathode Solution Hydranal Coulomat CG

2. Performance

Parameter Acceptance
Blank Rate <5 µg H₂O/min
Water Standard Recovery 98 – 102%
Sample Analysis Time 2 – 5 min