Water Purification System — Standard Operating Procedure

1. Daily Checks

  • [ ] Record resistivity reading (target: ≥18.2 MΩ·cm)
  • [ ] Verify TOC reading (<5 ppb)
  • [ ] Check RO pressure (4–6 bar)
  • [ ] Inspect for visible leaks
  • [ ] Verify dispense flow rate
  • [ ] Log all readings in water quality logbook

2. Usage Guidelines

Application Water Grade Notes
SPPS resin washing Type 2 Freshly dispensed
HPLC mobile phase Type 1 (18.2 MΩ·cm) Filter through 0.22 µm
Lyophilization diluent Type 1
LC-MS solvents Type 1 TOC <5 ppb critical

3. Routine Maintenance

Interval Task
Weekly Sanitize dispense point with 70% IPA
Monthly Replace 0.22 µm final filter
Quarterly Replace carbon filter and sediment filter
Semi-annually Replace UV lamp
Annually Replace RO membrane and ion exchange resins

4. Out-of-Specification Response

If resistivity falls below 18.0 MΩ·cm: 1. Stop dispensing water 2. Check system display for error codes 3. Flush system for 10 min at max flow 4. If persists, call maintenance 5. Document incident