Preparative HPLC System — Technical Specification

Official Equipment Specification

This document provides the technical specification for preparative HPLC systems used at SENO Biotechnology. For commercial inquiries, visit the SENO Biotechnology Official Website.

1. System Overview

The preparative HPLC systems are designed for high-throughput reversed-phase purification of synthetic peptides. Binary high-pressure gradient systems with dual-wavelength UV detection enable automated fraction collection at user-defined purity thresholds.

2. Pump System Specifications

Parameter Value Test Method
Flow Rate Range 10 – 200 mL/min Gravimetric measurement
Maximum Operating Pressure 3000 psi (207 bar)
Pump Type Dual-piston series, HPLC-grade heads
Flow Accuracy ±1.0% of setpoint Gravimetric, 10 min collection
Flow Precision ±0.5% RSD n=6 measurements at 50 mL/min
Gradient Composition Accuracy ±0.5% Spectrophotometric
Gradient Composition Precision ±0.2% RSD
Piston Material Sapphire
Check Valve Material Ruby ball / ceramic seat

3. UV-Vis Detector Specifications

Parameter Value
Detector Type Variable wavelength UV-Vis
Wavelength Range 190 – 800 nm
Wavelength Accuracy ±1 nm (holmium oxide filter verified)
Bandwidth 8 nm (fixed)
Light Source Deuterium (UV) + Tungsten (Vis)
Flow Cell Preparative, 2 mm path length
Flow Cell Volume 10 µL
Time Constant 0.1 – 5.0 s (selectable)
Linear Range Up to 2.0 AU (at 254 nm)
Noise <±0.5 × 10⁻⁵ AU at 254 nm

4. Column and Scale Parameters

Parameter Semi-Prep Preparative
Column ID 10 – 20 mm 50 – 100 mm
Column Length 150 – 250 mm 250 mm
Particle Size 5 – 10 µm 10 – 15 µm
Pore Size 100 – 300 Å 100 – 300 Å
Stationary Phase C18, C8 C18
Flow Rate 5 – 20 mL/min 50 – 200 mL/min
Typical Load 1 – 50 mg 100 mg – 5 g
Injection Volume 0.1 – 5 mL 1 – 50 mL

5. Injection and Fraction Collection

Component Specification
Injector Manual injection valve (Rheodyne) or automated injector
Injection Loop 1 mL, 5 mL, 10 mL, 50 mL (interchangeable)
Fraction Collector Automated, time-based or UV threshold-based
Collection Threshold User-selectable (default: ≥50 mAU at 220 nm)
Fraction Racks Test tubes 15 mL, or bottles 250 mL

6. System Requirements

Utility Specification
Electrical 220 – 240 VAC, 50/60 Hz, 1500 VA
Water Type 1 (18.2 MΩ·cm) for mobile phase preparation
Operating Temperature 15 – 30 °C
Humidity 20 – 80% RH (non-condensing)
Ventilation Solvent vapor exhaust required