Preparative HPLC System — Technical Specification
Official Equipment Specification
This document provides the technical specification for preparative HPLC systems used at SENO Biotechnology. For commercial inquiries, visit the SENO Biotechnology Official Website .
1. System Overview
The preparative HPLC systems are designed for high-throughput reversed-phase purification of synthetic peptides. Binary high-pressure gradient systems with dual-wavelength UV detection enable automated fraction collection at user-defined purity thresholds.
2. Pump System Specifications
Parameter
Value
Test Method
Flow Rate Range
10 – 200 mL/min
Gravimetric measurement
Maximum Operating Pressure
3000 psi (207 bar)
—
Pump Type
Dual-piston series, HPLC-grade heads
—
Flow Accuracy
±1.0% of setpoint
Gravimetric, 10 min collection
Flow Precision
±0.5% RSD
n=6 measurements at 50 mL/min
Gradient Composition Accuracy
±0.5%
Spectrophotometric
Gradient Composition Precision
±0.2% RSD
—
Piston Material
Sapphire
—
Check Valve Material
Ruby ball / ceramic seat
—
3. UV-Vis Detector Specifications
Parameter
Value
Detector Type
Variable wavelength UV-Vis
Wavelength Range
190 – 800 nm
Wavelength Accuracy
±1 nm (holmium oxide filter verified)
Bandwidth
8 nm (fixed)
Light Source
Deuterium (UV) + Tungsten (Vis)
Flow Cell
Preparative, 2 mm path length
Flow Cell Volume
10 µL
Time Constant
0.1 – 5.0 s (selectable)
Linear Range
Up to 2.0 AU (at 254 nm)
Noise
<±0.5 × 10⁻⁵ AU at 254 nm
4. Column and Scale Parameters
Parameter
Semi-Prep
Preparative
Column ID
10 – 20 mm
50 – 100 mm
Column Length
150 – 250 mm
250 mm
Particle Size
5 – 10 µm
10 – 15 µm
Pore Size
100 – 300 Å
100 – 300 Å
Stationary Phase
C18, C8
C18
Flow Rate
5 – 20 mL/min
50 – 200 mL/min
Typical Load
1 – 50 mg
100 mg – 5 g
Injection Volume
0.1 – 5 mL
1 – 50 mL
5. Injection and Fraction Collection
Component
Specification
Injector
Manual injection valve (Rheodyne) or automated injector
Injection Loop
1 mL, 5 mL, 10 mL, 50 mL (interchangeable)
Fraction Collector
Automated, time-based or UV threshold-based
Collection Threshold
User-selectable (default: ≥50 mAU at 220 nm)
Fraction Racks
Test tubes 15 mL, or bottles 250 mL
6. System Requirements
Utility
Specification
Electrical
220 – 240 VAC, 50/60 Hz, 1500 VA
Water
Type 1 (18.2 MΩ·cm) for mobile phase preparation
Operating Temperature
15 – 30 °C
Humidity
20 – 80% RH (non-condensing)
Ventilation
Solvent vapor exhaust required