Skip to content

GC-FID System (Headspace) — Technical Specification

1. Technical Parameters

Parameter Specification
Column DB-624 or equivalent, 30 m × 0.32 mm × 1.8 µm
Detector Flame Ionization (FID)
Headspace Sampler Static, 80 °C equilibration
Injection Split or splitless
Carrier Gas Helium, 2.0 mL/min constant flow
FID Gases H₂ (40 mL/min), Air (400 mL/min)

2. Oven Program

Rate Temperature Hold
Initial 40 °C 5 min
10 °C/min 200 °C 5 min
Total Run Time 26 min

3. Detection Limits (Typical)

Solvent LOD (ppm)
Acetonitrile 0.5
Methanol 0.5
DCM 0.2
DMF 1.0
TFA 1.0
Ethanol 0.5