Analytical HPLC System — Technical Specification

Official Equipment Specification

This document provides the technical specification for analytical HPLC systems used at SENO Biotechnology. For commercial inquiries, visit the SENO Biotechnology Official Website.

1. Technical Parameters

Parameter Specification
Flow Rate Range 0.1 – 5.0 mL/min
Maximum Operating Pressure 6000 psi (414 bar)
Solvent Delivery Quaternary or binary gradient
Degasser Online vacuum degassing (4 channels)
Injector Autosampler with sample cooling (4 – 40 °C)
Injection Volume 0.1 – 100 µL
Column Compartment Peltier-controlled, 10 – 85 °C

2. Detector Specifications

Parameter DAD (Diode Array) Specification
Wavelength Range 190 – 800 nm Full spectrum
Lamp Deuterium + Tungsten Automatic switch
Bandwidth 1 – 40 nm Adjustable
Data Rate Up to 100 Hz Real-time spectral acquisition
Noise <±0.5 × 10⁻⁵ AU At 254 nm

3. System Performance

Parameter Specification Test Method
Flow Precision <0.3% RSD Retention time repeatability
Injection Precision <0.5% RSD Area repeatability (n=6)
Gradient Accuracy ±0.5% Composition verification
Wavelength Accuracy ±1 nm Holmium oxide filter
Sensitivity (S/N) >1000 1 µL of 1 µg/mL caffeine

4. Column Specifications

Parameter Typical Value
Column Dimensions 150 × 4.6 mm, 250 × 4.6 mm
Particle Size 3 – 5 µm
Pore Size 100 – 300 Å
Stationary Phase C18, C8, C4, HILIC