Skip to content

Environmental Monitoring System — Technical Specification

1. System Parameters

Parameter Sensor Type Range Accuracy
Temperature PT100 RTD 0 – 50 °C ±0.3 °C
Relative Humidity Capacitive 10 – 90% RH ±2% RH
Differential Pressure Piezoresistive −50 to +50 Pa ±0.5 Pa
Particle Count Laser diode 0.5 – 25 µm ISO 21501-4

2. Monitoring Locations

Area Parameters Alarm Limits
Production Suites Temp, RH, DP 20 – 25 °C, 30 – 65% RH, ≥15 Pa
Analytical Labs Temp, RH 20 – 25 °C, 30 – 60% RH
Storage (4 °C) Temp 2 – 8 °C
Storage (−20 °C) Temp −25 to −15 °C