Preparative HPLC System — Standard Operating Procedure
1. Pre-Operation Safety
- Wear solvent-resistant gloves (Viton or nitrile) and safety goggles
- Verify adequate ventilation or connect to exhaust
- Confirm waste container has ≥50% empty capacity
- Inspect all tubing connections for leaks
2. Pre-Operation Checks
| Item |
Check |
Acceptable |
| Mobile phase A |
H₂O + 0.1% TFA (filtered 0.22 µm) |
Prepared fresh, labeled |
| Mobile phase B |
ACN + 0.1% TFA (filtered 0.22 µm) |
Prepared fresh, labeled |
| System pressure test |
Pump at 10 mL/min, 50% B, monitor 5 min |
<5% pressure fluctuation |
| UV lamp |
Warm up ≥15 min before use |
Baseline drift <0.5 mAU/min |
| Column condition |
Visual inspection (no end-fitting leakage) |
Dry exterior |
| Waste container |
Volume check |
≥50% empty |
| Fraction collector |
Tube/bottle positions confirmed |
Set collection mode |
3. System Equilibration
- Purge pump heads with mobile phase (5 mL/min each head, 2 min)
- Set flow rate to 10 mL/min at initial %B (typically 5% B)
- Allow column to equilibrate for 3 – 5 column volumes
- Monitor UV baseline at 220 nm (peptide bond) and 280 nm (aromatic)
- Accept equilibration when baseline drift <0.5 mAU/min over 2 min
- Record equilibration parameters in batch record
4. Sample Preparation
- Weigh crude peptide (record mass)
- Dissolve in mobile phase A or ACN/H₂O mixture (10 – 50 mg/mL typical)
- Filter through 0.45 µm PTFE syringe filter into clean vial
- Record sample concentration and volume
- Load sample into injection loop using Luer-lock syringe
5. Gradient Purification Method
| Step |
Time (min) |
Flow (mL/min) |
%B |
Curve |
Purpose |
| Equilibration |
0 – 5 |
50 |
5 |
— |
Ready column |
| Injection |
0 (event) |
50 |
5 |
— |
Inject sample |
| Gradient 1 |
0 – 5 |
50 |
5 → 20 |
Linear |
Wash non-retained |
| Gradient 2 |
5 – 45 |
50 |
20 → 55 |
Linear |
Elute target peptide |
| Gradient 3 |
45 – 50 |
50 |
55 → 95 |
Linear |
Column wash |
| Hold |
50 – 55 |
50 |
95 |
Isocratic |
Elute strongly retained |
| Re-equilibration |
55 – 65 |
50 |
95 → 5 |
Linear |
Return to start |
6. Fraction Collection Criteria
- Set collection threshold: ≥50 mAU at 220 nm (adjust for concentrated samples)
- Collect peak apex ±0.5 min for primary fraction
- Collect leading and trailing shoulders separately (may contain impurities)
- Label each fraction with:
- Batch number
- Fraction number
- Collection time window
- UV absorbance at apex
7. Post-Run Fraction Analysis
- Transfer 5 – 20 µL from each fraction to auto-sampler vial
- Analyze by analytical HPLC (see Analytical HPLC SOP)
- Pool fractions meeting purity threshold (≥95% for standard grade)
- Record pooled fraction volume, concentration (by UV), and total peptide mass
8. Column Cleaning and Storage
| Step |
Solvent |
Volume |
Purpose |
| Post-run wash |
100% ACN |
3 CV (50 mL/min) |
Remove hydrophobic impurities |
| Intermediate |
70% ACN / 30% H₂O |
3 CV |
Transition |
| Storage |
80% ACN / 20% H₂O |
— |
Prevent microbial growth |