Lyophilizer — Technical Specification

Official Equipment Specification

This document provides the technical specification for lyophilization systems used at SENO Biotechnology. For commercial inquiries, visit the SENO Biotechnology Official Website.

1. Technical Parameters

Parameter Specification Notes
Condenser Temperature −80 °C Two-stage cascade refrigeration, CFC-free
Shelf Temperature Range −40 °C to +60 °C Silicone oil circulation
Temperature Control Accuracy ±1.0 °C across all shelves Uniformity verified by 3-point mapping
Ultimate Vacuum <0.01 mbar Dry scroll or oil-sealed rotary vane pump
Vacuum Control Range 0.01 – 1.0 mbar Pirani gauge, adjustable setpoint
Shelf Material 316L Stainless Steel Ra ≤0.8 µm surface finish
Shelf Flatness ±0.5 mm For uniform heat transfer
Number of Shelves 4 – 8 (model-dependent) Adjustable spacing
Total Shelf Area 0.5 – 5.0 m² Model-dependent
Ice Capacity 10 – 40 kg Before defrost required
Refrigerant R-404A / R-508B (cascade) CFC-free, low GWP alternatives available

2. System Components

Component Type Specification
Chamber 316L SS cylindrical Pressure-rated for full vacuum to 2 bar overpressure
Condenser Internal coil, −80 °C Direct expansion refrigeration
Refrigeration Two-stage cascade Stage 1: −40°C, Stage 2: −80°C
Vacuum Pump Dry scroll or oil-sealed rotary vane 10 – 40 m³/h displacement
Control System PLC with color touchscreen HMI Recipe programming, data logging, alarm management
Temperature Sensors PT100 RTD (shelf-mounted) ±0.1 °C accuracy
Pressure Sensor Pirani + Piezo (dual) 1 × 10⁻³ to 1000 mbar
Vent Filter 0.22 µm hydrophobic PTFE Sterile nitrogen backfill

3. Cycle Phase Parameters

Phase Target Temp Ramp Rate Hold Time Vacuum Setpoint Purpose
Loading 4 °C Variable Atmospheric Product placement
Freezing −40 °C 0.5 – 2.0 °C/min 2 – 4 h Atmospheric Complete solidification
Primary Drying −10 °C 0.3 – 0.5 °C/min 12 – 36 h 0.05 – 0.1 mbar Ice sublimation
Secondary Drying +25 °C 0.1 – 0.3 °C/min 4 – 8 h <0.02 mbar Bound water desorption
Backfill +25 °C 5 min N₂ to atmospheric Product removal

4. Acceptance Criteria

Parameter Criteria Method
Residual Moisture ≤1.0% (bulk), ≤0.5% (formulated) Karl Fischer coulometric titration
Cake Appearance Uniform, no collapse or melt-back Visual inspection
Product Temperature Within 2 °C of shelf temperature at end of each phase In-process PT100 probe
Vacuum Stability Stable <0.02 mbar for ≥1 h before termination Pirani gauge trend
Pressure Rise Test <0.01 mbar/min Valve isolation test

5. Maintenance Schedule

Frequency Activity
After each cycle Chamber wipe-down with 70% IPA; inspect door seal
Weekly Condenser defrost (manual or automatic); drain condensate
Monthly Vacuum pump oil level check (if oil-sealed); inlet filter inspection
Quarterly Temperature sensor calibration (PT100 vs certified reference, ±0.5 °C)
Semi-annually Vacuum pump oil change (oil-sealed) or bearing inspection (dry)
Annually Refrigerant pressure check; vacuum leak test; full PM