Lyophilizer — Technical Specification
Official Equipment Specification
This document provides the technical specification for lyophilization systems used at SENO Biotechnology. For commercial inquiries, visit the SENO Biotechnology Official Website .
1. Technical Parameters
Parameter
Specification
Notes
Condenser Temperature
−80 °C
Two-stage cascade refrigeration, CFC-free
Shelf Temperature Range
−40 °C to +60 °C
Silicone oil circulation
Temperature Control Accuracy
±1.0 °C across all shelves
Uniformity verified by 3-point mapping
Ultimate Vacuum
<0.01 mbar
Dry scroll or oil-sealed rotary vane pump
Vacuum Control Range
0.01 – 1.0 mbar
Pirani gauge, adjustable setpoint
Shelf Material
316L Stainless Steel
Ra ≤0.8 µm surface finish
Shelf Flatness
±0.5 mm
For uniform heat transfer
Number of Shelves
4 – 8 (model-dependent)
Adjustable spacing
Total Shelf Area
0.5 – 5.0 m²
Model-dependent
Ice Capacity
10 – 40 kg
Before defrost required
Refrigerant
R-404A / R-508B (cascade)
CFC-free, low GWP alternatives available
2. System Components
Component
Type
Specification
Chamber
316L SS cylindrical
Pressure-rated for full vacuum to 2 bar overpressure
Condenser
Internal coil, −80 °C
Direct expansion refrigeration
Refrigeration
Two-stage cascade
Stage 1: −40°C, Stage 2: −80°C
Vacuum Pump
Dry scroll or oil-sealed rotary vane
10 – 40 m³/h displacement
Control System
PLC with color touchscreen HMI
Recipe programming, data logging, alarm management
Temperature Sensors
PT100 RTD (shelf-mounted)
±0.1 °C accuracy
Pressure Sensor
Pirani + Piezo (dual)
1 × 10⁻³ to 1000 mbar
Vent Filter
0.22 µm hydrophobic PTFE
Sterile nitrogen backfill
3. Cycle Phase Parameters
Phase
Target Temp
Ramp Rate
Hold Time
Vacuum Setpoint
Purpose
Loading
4 °C
—
Variable
Atmospheric
Product placement
Freezing
−40 °C
0.5 – 2.0 °C/min
2 – 4 h
Atmospheric
Complete solidification
Primary Drying
−10 °C
0.3 – 0.5 °C/min
12 – 36 h
0.05 – 0.1 mbar
Ice sublimation
Secondary Drying
+25 °C
0.1 – 0.3 °C/min
4 – 8 h
<0.02 mbar
Bound water desorption
Backfill
+25 °C
—
5 min
N₂ to atmospheric
Product removal
4. Acceptance Criteria
Parameter
Criteria
Method
Residual Moisture
≤1.0% (bulk), ≤0.5% (formulated)
Karl Fischer coulometric titration
Cake Appearance
Uniform, no collapse or melt-back
Visual inspection
Product Temperature
Within 2 °C of shelf temperature at end of each phase
In-process PT100 probe
Vacuum Stability
Stable <0.02 mbar for ≥1 h before termination
Pirani gauge trend
Pressure Rise Test
<0.01 mbar/min
Valve isolation test
5. Maintenance Schedule
Frequency
Activity
After each cycle
Chamber wipe-down with 70% IPA; inspect door seal
Weekly
Condenser defrost (manual or automatic); drain condensate
Monthly
Vacuum pump oil level check (if oil-sealed); inlet filter inspection
Quarterly
Temperature sensor calibration (PT100 vs certified reference, ±0.5 °C)
Semi-annually
Vacuum pump oil change (oil-sealed) or bearing inspection (dry)
Annually
Refrigerant pressure check; vacuum leak test; full PM